Then switch to »standby« (see page 48) and call up
the leakage test so that the new compliance can be
calculated.
Invoking the leakage test
1 Press »TEST« on the ventilator for at least three
seconds.
2 Display in dialogue field on ventilator:
Leaktest IPPV
followed by:
Leak IPPV = xx mL
The calculated leakage and compliance values are
saved and displayed on the data screen.
The manual ventilation bag and its hose are not
included in the test!
The system switches back to »standby« after the test.

Anaesthesia ventilation with the Kuhn system

If the machine does not have an external fresh gas outlet
(optional - see page 38), anaesthesia ventilation with the
Kuhn system is performed as follows.:
Prepare the Kuhn set in accordance with separate
Instructions for Use –
connect anaesthetic gas scavenging hose
(see page 92).
3 Press »MAN/SPONT« on ventilator for at least one
second –
4 set pressure limiting valve APL to »MAN«.
Connect inspiration hose via
5 connecting sleeve –
6 Expiration connector remains open – Symbol:
7 Seal manual ventilation connector
with cap.
The pressure indicated on the monitor is not identical
with the actual airway pressure. Reason: higher flow
resistance of the fresh gas hose in the Kuhn system.
The higher the fresh gas flow, the greater the difference.
8 The anaesthetic and O
monitored by means of intake measurement with the
monitor in HLM mode.
Symbol:
Symbol:
concentration must be
2
Anaesthesia ventilation with the Kuhn system
2
4
3
8
5
6
7
Operation
Paediatric use
1
43