C-Lock™ ECG synchronization (optional)
Prerequisite for the
ECG synchronization signal ......: Pos. pulse with > 4.5 V and > 10 ms duration to drive 2 mA.
Max. permitted signal delay
in relation to the current
QRS complex ...........................: 40 ms
Signal isolation from
remaining electronics ................: Dielectric strength > 4 kV
Socket for 2-pin pawl-type
connector, dia. 3.5 ....................: Pin assignment:

Operating parameters

Noise emission ..................................... : Max. 56 dB(A) (as determined in a free field over a reflecting surface)
Protection classes:
SpO
sensor ....................................... : Type BF
2
Electromagnetic compatibility ............... : Tested to EN 60601-1-2
Classification ...................................... : Class II b
as per Directive 93/42/EEC
Annex IX
UMDNS-Code ..................................... : 10-134
Universal Medical Device
Nomenclature System
Technical documentation available on request
signal
ground
m
electrically isolated from protective earth conductor
Technical data
Monitor
Operating parameters
121