Dismantling parts

T-piece of measured gas hose
1 Remove filter.
Inspiratory O
sensor
2
Must not be disinfected in liquid or autoclaved!
Wipe any dirt off housing or lead with damp
disposable cloth.
Unscrew O
sensor and remove O
2
touch wire mesh of O
2
water-repellent property and measurement becomes
inaccurate.
Dirt on wire mesh of sensor cell is simply wiped off
with a disposable cloth moistened with distilled water.
Temperature sensor (breathing gas)
Draw temperature sensor out of Y-piece, remove
Y-piece from ventilation hoses.
Prise lead out of hose clips.
Pull connector out of socket.
Wipe dirt off with damp disposable cloth.
Flow sensor
The flow sensor cannot be disinfected or sterilized.
Contamination by infectious patients: dispose of as
directed!
If contamination is unlikely, the flow sensor can be
reused as long as it can be calibrated correctly.
(if used)
(if used)
cell. Do not
2
cell, otherwise it loses its
(optional)
T-piece of measured gas hose
Temperature sensor, Flow sensor
1
Care
Dismantling parts
Inspiratory O
sensor
2
81