Configuration During Operation - Dräger Cato Instructions For Use Manual

Anaesthetic workstation
Table of Contents

Advertisement

Monitor functions
Softkeys

Configuration during operation

Configuration during operation
The menu displayed during operation shows the functions
required for measuring mode. The functions are selected
as described for other menus. The settings made here only
apply for the present operation.
Press the softkey »config.«
The configuration menu is displayed. The values
shown against a grey background currently apply.
Turn the rotary control to move the cursor frame as
in the standby configuration menu. Refer to the
description of that menu for further details (page 51
onwards).
set
– pulse tone
For setting the volume.
– alarm tone
For setting the volume.
– mode
Change over between adult and
neonate modes.
– parameters
For setting the parameters, see
page 53.
– record
For determining which event
triggers an entry in the record list or
printout on the printer, see page 53.
– activate defaults Activates the configured default
settings (see page 51 onwards).
calibrating
– calibration
O
2
others.
warning
– default
Activates all default alarm limits
configured in »standby«.
– CO
Activates/deactivates all CO
2
alarms (including »Apnoea«).
– SpO
Activates/deactivates all SpO
2
alarms.
– HLM mode
Activates/deactivates the heart/lung
machine monitoring mode. See
page 68.
– P
»on«
Ventilator pressure limitation
max
generates a caution message.
»off«
Ventilator pressure limitation
generates an advisory message.
The »off« setting is recommended
for ventilation with intentional
pressure limitation.
– anaesth. gas
see »Configuring monitor functions
in « standby »« on page 56.
64
and flow sensor, as well as
2
Man./spont. alarm limits
CO
40
20
0
PAW
20
0
IPPV alarm limits
anaesth. gas
warning
Halothane
default
Enflurane
Isoflurane
CO
on
2
Sevoflurane
SpO
on
2
Desflurane
HLM
on
no
anaesth. gas
P
max
on
2
2
98
SpO
2
38
etCO
2
6.0
MV
Fi
29
O
18
2
0.8
Hal.
70
N
O
2
calibrating
set
O
-sensor
2
pulse to.
0 1 2 3 4
21 Vol.%
alarm tone
off
mode
flow-sensor
parameters
off
record
activate defaults
more
off
off
Select and confirm !
67
alarm
limits
CO
2
al. off
10
freq
alarm
info
Fet
list
25
curve
0.6
68
config.
5
6 7 8 9
1 2 3
4
5 6 7 8 9
adult
Neo.