Breathing System; Soda Lime; Emergency Ventilation Bag - Dräger Cato Instructions For Use Manual

Anaesthetic workstation
Table of Contents

Advertisement

Breathing system

Lift off table top:
1 Lever set to the position shown in black.
The following are complete:
2 Hose with manual ventilation bag
Symbol:
(Connected from below – not shown)
– Correct breathing hoses installed.
(Adult or infant hoses)
3 Pressure measuring hose with filter connected.
4 Measured gas return hose
Symbol:
connected.
5 Fresh gas hose plugged in.
(Connected from below – not shown)
6 Valve discs inserted.
– Pressure limiting valve (APL) present.
– Inspiratory microbial filter.
Symbol:
– Expiratory microbial filter.
Symbol:

Soda lime

7 Lime has not noticeably changed colour (purple).
– Filling level adequate (up to the mark).
– Soda lime container is securely tightened – up to the
end stop (clockwise).

Emergency ventilation bag

(not shown)
– Bag is complete with mask and hung from the side
of the Cato.
– Bag functions correctly.
Check and tick off.
Preparing for use
Checking readiness for operation with checklist
Breathing system, Soda lime
Emergency ventilation bag
1
2
5
7
3
6
4
APL
Page 23