Dräger Cato Instructions For Use Manual page 33

Anaesthetic workstation
Table of Contents

Advertisement

1 Display:
Leak IPPV = xxx mL
If greater than 175 mL/min:
Repair leak, particularly for low-flow, or continue if
not problematical –
Confirm:
2 Press rotary control on ventilator!
1 Display with test block No. 9:
Leaktest MAN
1 Display:
Self-test
Wait – approx. 1 minute!
A tick ( ) appears on the screen when the ventilator test
has been completed successfully.
The prompt »Confirm« appears.
Press rotary control on monitor.
The cursor jumps to the arrow symbol on the
configuration menu and the vertical pressure indication
disappears.
Press the rotary control on the monitor again to
confirm and exit the configuration menu. Configuration
is now complete.
Tick off the ventilator self-test and monitor self-test on
the checklist.
Checking readiness for operation with checklist
1
Standby / Configuration
anaesth. gas
warning
calibrating
Halothane
default
O
-sensor
Enflurane
2
21 Vol.%
Isoflurane
flow sensor
Sevoflurane
Ventilator
start up test
Desflurane
more
no
anaesth. gas
Preparing for use
Ventilator start-up test
1
2
Alarms inactive!
defaults
pulse to.
0 1 2 3 4
5
6 7 8 9
alarm tone
1 2 3
4
5 6 7 8 9
mode
adult
Neo.
parameters
record
interfaces
alarm limits
curves
basic configuration
33