Dräger Cato Instructions For Use Manual page 93

Anaesthetic workstation
Table of Contents

Advertisement

Fit flow sensor
1 Unscrew expiration nozzle.
2 Slide flow sensor into housing.
1 Screw expiration nozzle back into place.
Fit microbial filter
Fit a microbial filter with ISO nozzle on both the
inspiration and expiration legs.
Fit temperature sensor
(Optional) with: Y-piece M 30 543 and
hose clips 84 04 047.
The T-piece must be used for anaesthetic gas measure-
ment in this case!
3 Insert temperature sensor 84 05 371 as far as
possible into the hole in the Y-piece.
After securing the Y-piece, align the sensor to face
upwards so that it remains free from condensation.
4 Route the sensor lead back to the machine with hose
clips along the inspiration hose of the anaesthetic
system.
Install secretion aspirator
Position the secretion aspirator on the mount on the
frame with two secretion collecting vessels.
5 Install hose between negative pressure outlet and
secretion collecting vessels.
6 Connect the gas outlet on the Cato to the ejector.
7 Plug the aspiration hose with handpiece onto the inlet.
Secure the aspiration hose in the clamp.
Secretion aspirator may only be used in
[
»MAN/SPONT« mode or with disconnected
Y-piece.
Fit flow sensor and microbial filter
2
1
3
6
5
Care
Reassembling system
Fit temperature sensor
Install secretion aspirator
4
4
7
93