Dräger Cato Instructions For Use Manual page 32

Anaesthetic workstation
Table of Contents

Advertisement

Preparing for use
Checking readiness for operation with checklist
Ventilator start-up test
The dialogue continues on the ventilator:
1 Test block No. 3 and display:
Fresh gas occl?
3 Close all three delivery valves!
Confirm:
2 Press rotary control on ventilator!
1 Display:
APL = 30 mbar?
4 Set pressure limiting valve to »MAN« and 30 mbar!
Confirm:
2 Press rotary control on ventilator!
1 Display:
Y-piece open?
5 Remove Y-piece from cone or exercise thorax!
Confirm:
2 Press rotary control on ventilator!
1 Display:
Self-test
1 Display:
Ypiece occluded?
5 Plug Y-piece onto cone!
(The sampling line is connected to the Y-piece and
the water trap.)
Confirm:
2 Press rotary control on ventilator!
1 Display with test block No. 4:
Self-test
The test then continues automatically, block by block:
wait – approx. 1 minute!
1 Display:
Leaktest IPPV
32
1
1
2
3
5
4