Oxisensor I-20 - Dräger Cato Instructions For Use Manual

Anaesthetic workstation
Table of Contents

Advertisement

Oxisensor I-20

Adhesive sensor for short and long-term monitoring
of patients with limited mobility and weighing between
3 and 15 kg.
Peel the protective film off the adhesive strip.
Place the sensor underneath the big toe so that the
dotted line corresponds to the inner edge of the toe
and the marking is in the middle of the toe.
Wrap the sensor round the toe so that the other
marking rests exactly on top of the toenail.
Secure the sensor lead to the foot with the additional
adhesive tape.
Reusing the sensor:
The sensor can be reused if the tape is still tacky.
Adhesion is improved by small additional adhesive spots.
Hold the adhesive spots by the blue tags, peel them
off the backing paper and remove the protective film.
Affix one spot to the middle of each optical element.
Position the sensor as described above.
Other measuring point:
The sensor is preferably applied to the big toe because it
moves less than the patient's hand. If the big toe is not
available, however, the sensor can also be applied to the
thumb.
Peel off the protective film on the adhesive strip.
Position the sensor under the thumb so that the
dotted line corresponds with the edge of the thumb
and the marking is in the middle of the thumb.
Wrap the sensor round the thumb so that the other
marking lies exactly on top of the thumb nail.
Secure the sensor lead to the hand with the additional
adhesive tape.
Care
Reassembling system
SpO
measurement
2
Applying the sensors
99