Manual Calibration Of The Flow Sensor - Dräger Cato Instructions For Use Manual

Anaesthetic workstation
Table of Contents

Advertisement

Preparing for use
Checking readiness for operation with checklist

Manual calibration of the flow sensor

Manual calibration of the flow sensor
– The flow sensor can be calibrated by hand while O
calibration is still in progress, either in the machine or
after being removed from the machine.
Without removal:
On the anaesthetic unit:
1 Set AIR/N
O selector to »AIR«.
2
2 Close the delivery valves for O
delivery valve for air and thoroughly flush the
breathing system with air.
3 Close the delivery valve for air.
Press the rotary control on the monitor to start
calibration.
The question mark (?) disappears and is replaced by
a timer icon (
) on the screen.
u
A tick (
) appears on the screen when calibration
has been completed correctly.
Tick off in checklist.
The cursor frame has automatically jumped to
»Ventilator start-up test«.
With removal:
4 Remove flow sensor:
– Unscrew expiration nozzle
– Remove flow sensor
– Briefly swing it to flush with ambient air
– Hold it horizontally with the cable connection pointing
downwards (calibration in installation position
improves the measuring accuracy)
– Seal off one or both sides as shown on the right,
preferably with your thumb or palm.
Press the rotary control on the monitor to start
calibration; continue as above.
4 Replace flow sensor!