Technical data
General
Power supply
Ambient conditions
Where tolerances are given in percent and as absolute values,
the higher value applies in each case!
General
Identification .............................: The Serial No. and Article No. are located on the rating plate (nameplate) on the
Weight .....................................: Cato basic machine approx. 140 kg
Dimensions...............................: See dimensional drawings on page 114
Power supply
Rated operating voltages ..........: depending on
Power input .............................: Max. 280 W (without auxiliary power sockets and without accessories)
Fuses .......................................: The fuses in the power adapter and master switch on contactor are only accessible
Auxiliary mains sockets .............: Do not connect any HF surgery devices to the auxiliary mains sockets!
Ambient conditions
Temperature .............................: + 15 to + 35 °C
Humidity ...................................: 20 to 80 %, no condensation (operation)
Air pressure..............................: 80 to 106 kPa
116
back of the machine. The Article No. also specifies the equipment level of the
machine.
Cato with two Vapors and monitor approx. 180 kg
100 V
AC
power adapter 120 V
AC
127 V
AC
230 V
AC
240 V
AC
Conversion only by specially trained personnel.
to qualified personnel with special tools.
Auxiliary power sockets (3x): 2 fuses installed per socket (2A DIN 41 622,
230/240 V
)
AC
The connection of devices to the auxiliary mains socket may result in increased
risk of electric shock in the event of earthing failure!
Auxiliary mains sockets
– 20 to + 50 °C
0 to 98 %, no condensation (storage / transport)
50 to 110 kPa
+10 %, –15 %, 50/60 Hz
±10 %, 50/60 Hz
±10 %, 50/60 Hz
±10 %, 50/60 Hz
±10 %, 50/60 Hz
(operation)
(storage / transport)
(operation)
(storage / transport)