Dräger Cato Instructions For Use Manual page 125

Anaesthetic workstation
Table of Contents

Advertisement

Flow chart of self-test in ventilator (cont.)
Test blocks:
7
F
A Fresh gas occl?
8
F
A
Leakage >10L/min
9
F
A BAG/Fr.gas hose?
F
A Fresh gas occl?
F
A BreathSys.defect
10
End of self-test
Control pressure test
Valve and roller diaphragm
test
> 5 L/min
Subsystem 1 / 2 (IPPV)
Leak < 1.5 L/min at
30 mbar?
I
Leaktest IPPV
Leak IPPV=
mL
A
< 175
> 175
I
Leaktest MAN.
Subsystem 3
(breathing bag / fresh gas line)
Leak < 300 mL/min
at 30 mbar
Breathing system
Valve function test
I
Standby
A
pneum.Interface?
F
A
Service Nr. 118
A
Service Nr. 119
A
Service Nr. 115
A
Valvedisk defect
F
A
Diaphragm defect
F
A
Subsyst.1/2 leak
F
A
Insp. Valve ?
F
A
Subsystem 1 leak
F
A
Subsystem 2 leak
F
A
Insp. Valve ?
F
A
Subsystem 1 leak
F
A
Subsystem 2 leak
F
A
Subsystem 3 leak
A
Leak =
mL/min
F
A
Fresh gas occl?
F
A
Insp. Valve ?
F
A
APL = 30 mbar ?
F
APL → MAN. ?
A
F
A
Subsyst.1/2 leak
F
Descriptions
Flow chart of self-test
Leak eleminated?
A
Leak accepted ?
Leak eleminated?
A
Leak accepted ?
125

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents