Oxisensor R-15 - Dräger Cato Instructions For Use Manual

Anaesthetic workstation
Table of Contents

Advertisement

Care
Reassembling system
SpO
measurement
2
Applying the sensors

Oxisensor R-15

Disposable adhesive sensor for short and long-term
monitoring of inactive patients weighing more than
50 kg. Preferably used for patients possibly suffering
from severe vasoconstriction or poor circulation.
Clean the bridge of the nose with the liquid in the
enclosed ampoule – keep away from the eyes.
Peel off the protective film on the sensor.
Align the sensor symmetrically on the bridge of the
nose: the two symbols should lie on the bone/cartilage
boundary.
Press the sensor firmly into place and hold for
10 seconds to ensure good adhesion.
The R-15 sensor must not be used on patients with
nasal intubation or a mask.
100