Medical gas supply

Pneumatic connections...............: Piped medical gas supply – O
Required pressures (piped supply): O
Maximum gas consumption.........: O
O
shortage signal .....................: Whistling signal tone.
2

Breathing system

Gas volume ................................: Approx. 3 L total enclosed gas volume
Compliance................................: Approx. 3 mL/mbar
Absorber volume ........................: 1.5 L soda lime Drägersorb 800
Pressure relief ............................: 90 ± 10 mbar
Pressure limiting valve ................: Variable between 5 and 70 mbar, ± 15% of set value
Patient connections ....................: ISO cone 22 mm ;
Anaesthetic gas scavenging .......: Nominal socket diameter 27 mm or ISO cone 30 mm ; (= EN 1281-1)
Leakage ventilation system .........: < 120 mL/min at 30 mbar
Maximum pressure in
breathing system ........................: 10 to 80 mbar ± 20%
Expiratory and
inspiratory resistance..................: ≤4 mbar with a flow of 60 L/min (according to EN 740)

Ventilator

Drive gas consumption ...............: 2 L/min O
Tidal volume (V
) ........................: < 20 mL
T
Frequency (f
) ........................: 6 to 80 per minute
IPPV
Minute volume (*
).....................: to 25 L/min
E
T
:T
...........................................: 1:3 to 2:1
I
E
T
:T
.........................................: 0 to 60 %
IP
I
Inspiratory flow *
max...............: max. 75 L/min
E
) .........................: 3 to 80 per minute ± 5 % (f
Frequency (f
IMV
PEEP .........................................: 0; 2 to 20 mbar
Trigger pressure.........................: – 1 mbar
Spare gas cylinders 11 L or 3 L for O
Anaesthetic gas scavenging
: 2.7 to 5.5 bar
2
N
O : 2.7 to 5.5 bar
2
AIR : 2.7 to 5.5 bar
: 100 L/min at 5 bar ± 10%, including 50 L/min at 5 bar ± 10% for O
2
O : 20 L/min at 5 bar ± 10%
N
2
AIR : 50 L/min at 5 bar ± 10%
Triggered: at latest when O
Stops: at latest when O
pressure rises above 260 kPa (2.6 bar)
2
Approx. 4 mL/mbar with absorber
Binds approx. 150 L CO
2
or +3 mbar, but at least 10 mbar over PEEP
or AIR in operation
2
0 L/min in »standby« mode
± 30 % or ± 6 mL
± 10 % or ± 10 mL
20 to 100 mL
± 5 % or ± 15 mL
100 to 1400 mL
± 5 %
± 5 %
± 10 % or ± 2 mbar (P
± 0.5 mbar (invariable)
, N
O, AIR
2
2
and N
O (optional)
2
2
pressure drops below 170 kPa (1.7 bar)
2
, sufficient for approx. 6 hours of operation
< f
)
IMV
IPPV
– PEEP > 10 mbar)
max
Technical data
Medical gas supply
Breathing system
Ventilator
flush
2
117