Parameters
The settings for the following parameters are defined via
this menu:
– SpO
-measurement
2
– sidestream
measurement O
2
– sample rate
– CO
units
2
Record
This menu is used to define which event triggers an
automatic entry in the record list or printout on the
on-line printer:
– time intervall
Entry upon expiry of a fixed time
interval in minutes.
– NiBP started
Entry after every NiBP
measurement with new results.
– warning started Entry is made whenever a warning
has been triggered.
– caution started
Entry is made whenever a caution
has been triggered.
Interfaces
The interface for data transfer (Dräger MEDIBUS RS 232)
and the printer port are configured here:
– baud rate
Transmission rate (can be varied,
see Instructions for Use of the device
to be connected)
– parity
Cannot be changed in MEDIBUS and
is merely displayed for information.
– data bits
Cannot be changed in MEDIBUS and
is merely displayed for information.
– stop bits
Cannot be changed in MEDIBUS and
is merely displayed for information.
– protocol
Allows the printer port to be used as
a second MEDIBUS interface.
on • off
on • off
60 • 200 mL/min
mmHg • kPa • % by volume
Configuring monitor functions in »standby«
Standby / Configuration
anaesth. gas
warning
calibrating
SpO
- measure.
on
off
2
sidestream
measurement O
on
off
2
sample rate
60
200
mL/min
CO
units
mmHg
kPa
Vol%
2
Standby / Configuration
anaesth. gas
warning
calibrating
list entry
time
1
2
5
10
intervall
NiBP started
on
off
warning started
on
off
caution started
on
off
printer entry
time
1
2
5
10
intervall
NiBP started
on
off
warning started
on
off
caution started
on
off
Standby / Configuration
anaesth. gas
warning
calibrating
RS 232 (Medibus) :
baud rate
1.2
9.6
parity
even
data bits
8
stop bits
1
record (printer)
protocol
printer
Medibus
baud rate
1.2 2.4 4.8
9.6
parity
odd
even
data bits
7
8
stop bits
1
2
Monitor functions

Parameters, Record, Interfaces

Alarms inactive!
defaults
pulse to.
0 1 2 3 4
5
6 7 8 9
alarm tone
1 2 3
4
5 6 7 8 9
mode
adult
Neo.
parameters
record
interfaces
alarm limits
curves
basic configuration
Menu for configuration
of measurement-parameters.
Alarms inactive!
defaults
pulse tone
0 1 2 3 4
5
6 7 8 9
off
alarm tone
1 2 3
4
5 6 7 8 9
mode
adult
Neo.
parameters
record
interfaces
alarm limits
off
curves
basic configuration
Menu for configuration
of list and print entries.
Alarms inactive!
defaults
pulse tone
0 1 2 3 4
5
6 7 8 9
alarm tone
1 2 3
4
5 6 7 8 9
mode
adult
Neo.
parameters
record
interfaces
alarm limits
curves
19.2
basic configuration
none
Menu for configuration
of Medibus and printer
ports.