Procleix TIGRIS Operator's Manual page 99

Table of Contents

Advertisement

c) Code 39
d) Interleaved 2 of 5 (Do NOT enable—this barcode symbology has not been
validated for samples on the PROCLEIX
e) Code 128
f)
AGC Sampling Mode - Controls the Automatic Gain Control (AGC). Unless
otherwise instructed by Novartis Diagnostics personnel, leave this setting as
Continuous.
g) Good Decode Reads - Specifies the number of identical reads that the scanner
must obtain from a barcode before reporting the sample barcode ID. The number of
reads can be set from 6 to 31, and the default is 6.
Configuring the Hand-held Barcode Scanner
The following instructions for configuring the hand-held barcode scanner are for use with
PROCLEIX TIGRIS System Software version 3.6 or higher. Before starting the procedure,
identify the model number of the hand-held barcode scanner to be configured. A metal plate
below the hand-held barcode scanner window indicates the model number (LS4208 or
LS4000i).
Configuring the Symbol LS4208 Hand-held Barcode Scanner
Refer to the following instructions to configure the Symbol LS4208 hand-held barcode scanner.
Note: Code 128, Code 39 (no check digit), and Interleaved 2 of 5 limited to 14
characters (I 2 of 5) are enabled by factory default.
1. Power on the system and computer workstation with the hand-held barcode scanner
installed.
2. Scan the following barcodes in this order, unless instructed otherwise:
®
PROCLEIX
TIGRIS System Operator's Manual
Volume I: Operating the Instrument
Allow Code 39 Barcodes - Enables interpretation of Code 39 barcodes.
Enable Check Digit - Enables verification of sample barcodes using a Modulo 43
Check Digit character. If this option is not selected, the scanner will assume that a
Check Digit is not present. If a barcode containing a Check Digit is scanned by the
system with this option disabled, the Check Digit will become part of the scanned
Sample ID. It will not be used to verify the scanned barcode.
Transmit Check Digit - Enables the check digit to be transmitted with the
barcode ID; it will be displayed in the Results Report.
Enable Large Intercharacter Gap - Controls the configuration frames. By default
this option is disabled, as it should only be enabled if Novartis Diagnostics
personnel determine that the barcodes being used have a large intercharacter
gap.
Allow Code 128 - Enables interpretation of non-ISBT Code 128 barcodes.
Note: It is recommended that the user contact Novartis Diagnostics Technical
Service to obtain further information before changing the Good Decode Reads
value.
S