Figure 2.10 Proper Ar Bottle Positioning - Procleix TIGRIS Operator's Manual

Table of Contents

Advertisement

d) Verify that the barcode for each bottle is visible through the slot in the carousel.
e) Lift the carousel by the knob at the top, supporting the carousel with the other hand.
With the support hand, remove the AR Carousel Cover and place the AR Carousel
gently into the slot in the AR Bay.
f)
Gently rotate the carousel until it locks into place.
The carousel is locked when a "click" is heard and a slight bit of resistance is felt as it
is rotated. The AR Carousel indicator on the Inventory Status screen should be green,
indicating that the carousel is seated properly.
g) Replace the AR Carousel Cover over the AR Carousel.
The notches on the AR Carousel Cover should fit snugly into the openings on the AR
Bay. Press down on one side of the AR Carousel Cover to verify that it is firmly seated
in the AR Bay and does not wobble back and forth.
h) Close the AR Door.
7. Load the Target Capture Reagent (TCR) into the TCR Carousel:
a) Lift open the Sample Door and TCR Door.
b) Slide the TCR bottle into the cup of the TCR Carousel that matches the number of
the quadrant of the AR Carousel where the other ARs were loaded.
c) Verify that the barcode of the bottle of TCR is visible through the slot in the TCR
cup. Rotate the bottle in the cup if necessary.
d) Close the TCR and Sample Doors.
8. Click Inventory.
The system reads the barcodes on the bottles and level-senses the reagents.
®
PROCLEIX
TIGRIS System Operator's Manual
Volume I: Operating the Instrument

Figure 2.10 Proper AR Bottle Positioning

S
I
ETTING UP THE
NSTRUMENT
Part # 902649EN Rev. A
2-23