Loading Assay Reagents; Working With Master Lots; Table 2.6 Ar Barcode Information - Procleix TIGRIS Operator's Manual

Table of Contents

Advertisement

S
2
ECTION

Loading Assay Reagents

Load Assay Reagents (ARs) onto the system after managing the Lower Bay inventory, priming
the system, and completing Daily Maintenance.
See the appropriate PROCLEIX

Working with Master Lots

The Master Lot barcode can be found on the Master Lot Sheet that is included in each AR kit.
The Master Lot ensures that only the reagents belonging to a specific Master Lot can be run
together on the PROCLEIX
Each Master Lot need only be entered once into the system. Use the Master Lot report to verify
that the barcodes on each new set of ARs matches the Master Lot information already entered.
Using the Master Lot Report
To view the Master Lot report:
1. Click inside the Inventory Status panel.
The Inventory Status screen displays.
2. On the right side of the Inventory Status screen, click View Master Lot Report....
A Select Master Lot Barcode window opens.
3. Select the Master Lot number from the Master Lot drop-down list.
A Master Lot Report window opens.
4. Verify that the individual lot numbers of each reagent on the Master Lot Report match the
lot numbers for the set.
The barcode on each AR bottle contains 21 digits. Table 2.6 describes the information
coded by the sets of digits that make up the barcode.
Information Type
Assay Type
Reagent Type
Lot #
Expiration date
Bottle ID
Adding a Master Lot
If a set of ARs belongs to a Master Lot that is not yet in the system database, the operator
must add the new Master Lot to the database before the set can be loaded.
1. In the Upper Bay Inventory panel on the Inventory Status screen, click New Lot. A New
Master Lot window opens.
2. Using the barcode scanner on the computer workstation, scan in the Master Lot barcode
on the Master Lot Sheet.
The Lot Number of the new Master Lot populates the Master Lot text box in the New
Master Lot window.
2-20
Part # 902649EN Rev. A
®
Assay package insert for reagent preparation instructions.
®
TIGRIS System.
# of
Digit Position
Digits
2
1-2
2
3-4
6
5-10
6
11-16
5
17-21

Table 2.6 AR Barcode Information

Example
03 = PROCLEIX ULTRIO
02 = Enzyme Reagent
123456
300606 = June 30, 2006
54321
®
PROCLEIX