Results Processing Errors - Procleix TIGRIS Operator's Manual

Table of Contents

Advertisement

A
A
PPENDIX
Code
Cause
TLH
A Left Heater temperature was out of
range.
TRH
A Right Heater temperature was out of
range.
TS
The ambient temperature in the Sample
Carousel area was out of range.
URM
MTU barcode error.
VVFS
The volume of the wTCR/Sample in the
MTU tube was higher or lower than
expected.
VVFT
The volume of the wTCR in the MTU tube
was higher or lower than expected.

Results Processing Errors

Note: The code descriptions in this table may not match those that appear in the legend
on the last page of the Results report. The descriptions listed here provide greater detail
regarding the specific cause of individual errors.
Code
Description
af
Post-kinetic index adjustment flasher
out of range for calibrator or control.
afc
Post-kinetic index adjustment specimen
flasher greater than cutoff.
ag
Post-kinetic index adjustment glower
out of range for calibrator or control.
agc
Post-kinetic index adjustment specimen
glower greater than cutoff.
cv
% cv out of range.
e
Data analysis error.
f
Failed run flag.
fa
Flasher average out of range.
fb
Front to back ratio out of range.
fl
Calibrator flasher out of range.
g
Calibrator glower out of range.
ga
Glower average out of range.
ic
Invalid sample became valid due to a
crossover.
A-4
Part # 902649EN Rev. A
Effect
This is a non-fatal error. All samples in
the affected MTU are marked invalid.
This is a non-fatal error. All samples in
the affected MTU are marked invalid.
Sample pipetting pauses, but assay
processing continues. No MTUs are
marked Invalid. Sample pipetting
continues when the gauge returns to the
ambient temperature.
The affected samples are marked Invalid.
This is a non-fatal error. All samples in
the affected MTU are marked Invalid.
This error does not count towards the
10% invalid criterion.
This is a non-fatal error. The affected
samples are marked Invalid.
This is a non-fatal error. The affected
samples are marked Invalid.
Effect
Test invalid.
Test invalid.
Test invalid.
Test invalid.
Test invalid.
Test invalid.
Test invalid.
Test invalid.
Test invalid.
Test invalid.
Test invalid.
Test invalid.
Test invalid.
®