Figure 6.2 Ar Report Example - Procleix TIGRIS Operator's Manual

Table of Contents

Advertisement

To generate the AR Report:
1. In the lower-right part of the Inventory Status screen, click View AR Report....
A Select Assay dialog box opens.
2. From the Assay drop-down list, select the appropriate assay.
The system generates an AR Report for all matched sets that have been scanned on the
system.
The first section of the Assay Reagent Report is information about matched sets of ARs that
are currently on-board the PROCLEIX TIGRIS System. This information is organized by
quadrant. For each matched set, the report lists the Master Lot barcode, Master Lot expiration
date, number of tests remaining, the open/thaw stability expiration date, and the amount of on-
board stability remaining. In addition, the report lists the bottle ID and lot number for each bottle
in the set.
Note: The system only tracks on-board stability while the software is running.
The second section of the AR report lists the same information for any unexpired matched sets
that are not currently on-board the PROCLEIX TIGRIS System.
The final section of the report lists the lot number, bottle ID, number of tests remaining, open/
thaw expiration date, and amount of on-board stability remaining for all unexpired Discriminatory
Probe Reagents that have been scanned into the system but are not currently on-board.
®
PROCLEIX
TIGRIS System Operator's Manual
Volume I: Operating the Instrument
Figure 6.2 AR Report Example (Page 2)
W
R
ORKING WITH
EPORTS
6-3
Part # 902649EN Rev. A