Code
Cause
RDR
RDV error.
RT
The barcode is the same as a sample
that has already been tested.
S
The barcode was manually scanned in
using the handheld scanner.
SID
"Fix Unreadable Barcodes before
Processing" was not enabled and a
carousel worklist had an unreadable
barcode. The system assigned a default
sample ID.
SPA
Tip error.
ST
Failure to pick a tip.
TAM
The Amplification Incubator temperature
was out of range.
TAN
The Anneal Incubator temperature was
out of range.
TAT
The Transition Incubator temperature
was out of range.
TC
A Chiller ramp temperature was out of
range.
TD
The ambient temperature in the
luminometer area was out of range.
TE
The ambient temperature in the
electronics cage was out of range.
TH
The HPA Incubator temperature was out
of range.
®
PROCLEIX
TIGRIS System Operator's Manual
Volume I: Operating the Instrument
Effect
This is a non-fatal error. All samples in
the affected MTU are marked invalid.
Note: All other reagent pipettor rinse
errors are fatal. The appropriate fatal
e r r o r m e s s a g e w i l l a p p e a r o n t h e
worklist status line of the report, but no
samples are flagged RDR.
The sample is not marked Invalid.
The sample is not marked Invalid.
The sample is not marked Invalid.
Sample pipetting is suspended. MTUs
that completed sample pipetting and
were removed from the Load Station
continue processing; results are
generated for these samples. Samples
that had not finished pipetting are flagged
as CAN.
This error does not count towards the
10% invalid criterion.
The affected samples are marked Invalid.
This error does not count towards the
10% invalid criterion.
This is a non-fatal error. All samples in
the affected MTU are marked invalid.
This is a non-fatal error. All samples in
the affected MTU are marked invalid.
This is a non-fatal error. All samples in
the affected MTU are marked invalid.
This is a non-fatal error. All samples in
the affected MTU are marked invalid.
Any MTU introduced to the luminometer
while the detect thermometer is too hot
will have its samples marked Invalid. All
samples in the MTU are flagged "TD,"
indicating that the temperature was out of
range.
A fatal error is raised and all samples of
all MTUs being processed are flagged
"TE" and marked Invalid.
This is a non-fatal error. All samples in
the affected MTU are marked invalid.
S
C
TATUS
ODES
A-3