Stopping A Run - Procleix TIGRIS Operator's Manual

Table of Contents

Advertisement

Further specimen manipulation should be performed according to laboratory SOPs (Standard
Operating Procedures).

Stopping a Run

Use the Assay Processing Status box in the Run Manager screen to stop an in-process worklist
one of two ways:
Stop Current - Click this button to stop sample pipetting in the current worklist highlighted
in the In-Process Worklists grid. The system stops pipetting that worklist and the
remaining unpipetted samples are marked CAN for Cancel. Any samples that have
already been pipetted continue processing. The system then begins processing the next
worklist scheduled to be pipetted.
Stop All - Click this button to stop all processing for all worklists in the In-Process
Worklists grid and Queued Worklist grid. All the results that were not released at the time
Stop All was clicked are marked NTA and are lost. Any samples that were not pipetted
are marked with both NTA and CAN.
All queued worklists that have not started pipetting are dequeued. Sample racks must be
unloaded.
Note: Clicking Stop All will result with a hardware reset of the analyzer. At this time, the
software will go to Data Access Only mode. After the reset has completed, the instrument
will perform an unload function to remove all MTUs from the system. This procedure could
take up to 2½ hours.
®
PROCLEIX
TIGRIS System Operator's Manual
Volume I: Operating the Instrument
R
S
UNNING
AMPLES
3-13
Part # 902649EN Rev. A