Figure 2.9 Mtu Waste Cover - Procleix TIGRIS Operator's Manual

Table of Contents

Advertisement

S
2
ECTION
4. Fit a new MTU Waste Bag into the MTU Waste Container.
5. Fold the top of the bag down around the outside of the MTU Waste Container.
Note: Press bag down firmly against sides of the MTU Waste Container to ensure
that there are no wrinkles or air pockets that might catch an MTU as it is
discarded.
6. Slide a new MTU Waste Deflector over the rear-right corner of the MTU Waste Bag and
Container.
The Waste Deflector helps to hold the MTU Waste Bag in place and keeps MTUs from
accumulating on one side of the container.
7. Fold the edges of the MTU Waste Cover down at the perforations and place it over the top
of the MTU Waste Container. Align the cover opening with the MTU Waste Deflector at the
rear of the MTU Waste Container. Ensure that the MTU Waste Cover graphic is visible
(facing up) when installing the MTU Waste Cover.
8. Slide the MTU Waste Container back into its position in the Waste Drawer.
9. Replace Liquid Waste Containers 1 and 2 into the Waste Drawer. Reconnect the correct
fittings to the caps.
2-18
Part # 902649EN Rev. A

Figure 2.9 MTU Waste Cover

®
PROCLEIX
TIGRIS System Operator's Manual
Volume I: Operating the Instrument