Return Times Panel

The Return Times panel estimates when the operator should return to the system to perform an
action for the current worklist. If the operator does not perform the action at that time, a time-out
period begins, and processing pauses. The operator must take care to complete the action
before the maximum time-out period is exceeded (about three hours under normal conditions).
Field
Load MTUs
Load Tips
Load Racks
Next Results Set
Auto Logoff
Note: If any magnetic wash stations are out of service, the return times will be estimates.
See Volume III, Error Handling for more information.
®
PROCLEIX
TIGRIS System Operator's Manual
Volume I: Operating the Instrument
Meaning of Flashing Red Icon
No ARs are loaded.
The Lower Bay fluids inventory is invalid, the system needs to be primed, or
there is no current fluid inventory.
Fluid inventory is needed or invalid.
One or more doors is unlocked or open.
The number of available pipette tips is less than 97 or the Tips count is not
displayed.
The number of available MTUs is less than six.

Table 2.2 Inventory Status Panel Icon Indicators

Meaning/Action Required
Load MTUs at the estimated time displayed. This return time is dashed out
if there are enough MTUs to complete the in-process worklist or if the MTU
queue is empty.
Load pipette tips at the estimated time displayed. This return time is dashed
out if there are enough tips to complete the in-process worklist.
Load additional sample racks at the estimated time displayed. This return
time is dashed out if all racks on the in-process worklist are currently
loaded.
A new set of completed worklist results will be available at the estimated
time displayed.
Unless the operator performs an action by the specified time, the system
will automatically log the operator off. If this occurs, the operator must log
back on before performing additional activities. A system administrator must
adjust or turn off the Auto Logoff setting. See "Configuring System Settings"
on page 5-4.

Table 2.3 Return Times Panel

S
I
ETTING UP THE
NSTRUMENT
Part # 902649EN Rev. A
2-3