Configuring The Archive Parameters - Procleix TIGRIS Operator's Manual

Table of Contents

Advertisement

S
5
ECTION
Positive Calibrator IC
Average
Blank
Blank
Master Lot Number
Master Lot Expiration
Date
Instrument serial #
Firmware Revision
Site Information
Type of tube (e.g.,
"Negative Calibrator")
Blank
Blank
Test ID
ADM Version Number
TIGRIS Software
Version Number
Example tube data line of PROCLEIX WNV Assay LIS File (comma delimited):
MAS5698,NonReactive,WNV,100002,2004-03-30
02:04:28,IUO,235623,Valid,0,0.00,1.33942,miked,90024,44947,1063.000,180048.333,1462805.0
00,180317.667,,,032504,2025-12-31 23:59:59,HC+1030,HC+E5.01 02/20/04,10210 Genetic
Center Dr,Specimen,,,100002-001357-12-1,1.6.0.10.001,1.6.0.9
All of the above information appears on a single line in the file.

Configuring the Archive Parameters

Use the Archive Configuration to set archive parameters, including the archive path and auto-
archive frequency. An operator must have "Maintenance - Software Setup" rights to configure
the archive parameters.
The following data is saved to the location specified by the archive path, which may be the
®
PROCLEIX
all log files
kinetic data
results reports
worklists
database archive
LIS files are not saved to this location. For information on LIS data location, see "Configuring
the LIS" on page 5-19.
5-32
Part # 902649EN Rev. A
Average RLU value for Positive Calibrator IC.
Reserved.
Reserved.
Master Lot information as entered on the Inventory
Status screen.
Master Lot Expiration Date information as entered in
Inventory Status screen.
Serial # of the instrument, with the prefix "HC+".
Firmware version, with the prefix "HC+".
Identifying site information.
Negative Calibrator, Positive Calibrator, Specimen,
Negative Control, Positive Control
Reserved.
Reserved.
Test ID in the format: WorklistID-RackBarcode-
PositionInRack-ReplicateNum.
ADM Version Number.
PROCLEIX TIGRIS System Software Version
Number.
Table 5.8 PROCLEIX
TIGRIS System computer hard drive, a CD-RW, or a USB flash memory drive:
®
WNV Assay LIS Files
®
PROCLEIX
254 (15
typical)
254 (20
typical)
254 (10
typical)
254 (10
typical)
254 (10
typical)
TIGRIS System Operator's Manual
Volume I: Operating the Instrument
20
0
0
6
19
512
0
0
18