Software Configuration; Table 5.2 Configuration Options - Procleix TIGRIS Operator's Manual

Table of Contents

Advertisement

Software Configuration

System configuration parameters are grouped together in the Options box on the Configuration
tab.
To enable an option, check the box next to it. The system automatically saves the changes.
Table 5.2 describes available options.
Option
Auto Logoff after no
activity for:__ minutes
Require Double Entry of
Sample IDs for Test
Requests
Scanning Barcode
Confirms Sample ID
Fix Unreadable Barcodes
before Processing*
Auto Start Pipetting after
Rack Validation
Display Tip Count
Place Controls every __
Samples
Require Operator Logon
When Worklist Queued
Flag Duplicate Samples
Enable Manually Created
Worklists
*If this option is not enabled, the instrument will not allow "Flag Duplicate Samples" to be enabled.
®
PROCLEIX
TIGRIS System Operator's Manual
Volume I: Operating the Instrument
Description
Automatically logs off the current operator after the
specified amount of time without activity.
Requires double entry for manual keyboard entry.
Not applicable to PROCLEIX Assays.
Allows the operator to fix unreadable barcodes before
processing. Prevents the system from automatically
assigning a Sample ID; all barcodes must be
successfully scanned or manually entered. If this
option is not enabled, the system automatically
assigns a Sample ID in the case of an unreadable
barcode.
The system automatically starts pipetting after racks
have been validated.
Displays the pipette tip count on the
Sets the number of specimens between PROCLEIX
TIGRIS Controls when control bracketing is being
used. See "Control Bracketing" on page 3-2 for more
information. The default setting is 172 specimens, but
this can be changed to any value between 1 and 492
(single analyte assay) or 1 and 487 (triple analyte
assay).
Requires the operator to re-enter a user ID and
password to queue a worklist.
Flags sample barcodes that have been run on a
previous or in-process worklist.
Not applicable to PROCLEIX Assays.

Table 5.2 Configuration Options

S
P
UPERVISOR
Required
Setting
No
requirement
Enabled
Disabled
Enabled
Enabled
Inventory screen.
Enabled
®
No
requirement
No
requirement
Enabled
Disabled
Part # 902649EN Rev. A
ROCEDURES
5-5

Advertisement

Table of Contents
loading

Table of Contents