Saving To A Usb Flash Memory Drive; Figure 4.2 Example Of Two Usb Ports - Procleix TIGRIS Operator's Manual

Table of Contents

Advertisement

S
4
ECTION
Note: When the system indicates that a write operation has been completed, wait at least
7 minutes before attempting to eject the CD-RW disk. After 7 minutes, check to see if the
LED light on the CD-RW drive is blinking. If the light is blinking, the system is still writing
data to the disk and the operator must wait another 7 minutes before attempting to eject
the CD-RW disk. If the light is not blinking, eject the CD-RW disk.

Saving to a USB Flash Memory Drive

Log files, kinetic data, results reports, worklists, and the database archive can be saved to a
USB flash memory drive with the following specifications:
SuperTalent Flash Inc. SUPERFLASH 512 MB USB 2.0 Flash Memory Drive Model
STF_UT_USB20, or equivalent.
No pre-loaded software (e.g. U3).
Minimum capacity of 512 MB.
Dedicated for use with PROCLEIX TIGRIS System files. Do not use this dedicated USB
flash memory drive with other file types.
Note: LIS files may not be saved to a USB flash memory drive.
To save data to a USB flash memory drive:
1. Confirm that the system is in Standby mode.
2. Locate a free USB port on the PROCLEIX
A port may be located on the front or back of the workstation, and there may be
multiple ports. An example of two ports is shown in Figure 4.2. Do not unplug existing
hardware from a port to accommodate the USB flash memory drive.
3. Firmly insert the USB flash memory drive into the USB port.
Note: If a "Found New Hardware Wizard" message appears on the screen, remove
the USB flash memory drive from the USB port and contact Novartis Diagnostics
Technical Service for further instructions.
4. From the Configuration tab on the Configuration screen, confirm that the archive path is
set to the E:\ drive.
5. Save the data (refer to appropriate heading later in this section according to the type of
data being saved).
Note: Do not remove the USB flash memory drive until file saving is complete.
Premature removal may lead to corruption of the memory drive or system
shutdown.
4-6
Part # 902649EN Rev. A

Figure 4.2 Example of Two USB Ports

®
TIGRIS System workstation.
®
PROCLEIX
TIGRIS System Operator's Manual
Volume I: Operating the Instrument

Advertisement

Table of Contents
loading