S
5
ECTION
Require Results Verification – requires that a worklist be verified before it is
transferred to the LIS.
Allow Transmit of Unreadable Barcodes – not applicable to PROCLEIX Assays.
Set LIS Export Delimiter – sets the preferred delimiter format for the export. Once
checked, the radio button for either the Comma or Tab delimiter format must be
selected in order to override the default delimiter.
6. Click OK.
7. Click Start in the LIS Control panel.
The LIS status changes from "Stopped" to "Idle"; LIS files can now be transmitted to the export
location.

LIS File Export Formats

LIS files are comma or tab-delimited files containing a Cyclic Redundancy Check (CRC) at the
end. The CRC can be used by middleware or laboratory information systems to ensure that the
file has not been altered.
If using language-specific installation (other than English) of PROCLEIX
Software, some values of the LIS output fields are translated. Contact Novartis Diagnostics
Technical Service for further information.
All tubes are reported in the file, including calibrators and invalids, with each line except the last
containing the data for one tube. The last line in the file contains the literal characters "[end]"
followed by an 8-digit CRC. There are no characters following the CRC—not even a line feed.
The CRC is always the last 8 characters in the file.
File structure (example):
<Data for tube 1>
<Data for tube 2>
.....
<Data for last tube>
[end]<crc>