Appendix A. Status Codes; Instrument Processing Errors - Procleix TIGRIS Operator's Manual

Table of Contents

Advertisement

A
A. S
PPENDIX
The tables in this appendix contain descriptions of the codes that appear on the Results Report if an
error occurs during assay processing.
Contact Novartis Diagnostics Technical Service for further information about error conditions.

Instrument Processing Errors

Code
Cause
APV
Timing error.
An operation was not completed.
A robot or magnetic wash aspirator was
in the incorrect position.
The incorrect reagent was added.
The operation returned a non-zero status.
APV
Movement error.
Hardware error.
Incorrect MTU location.
An unmatched APL end message was
generated.
CAN
The sample had not been pipetted when
an operator clicked Stop Current or Stop
All.
The sample had not been pipetted when
a fatal error occurred, stopping
processing.
CLT
A clot was detected in the sample during
aspiration.
FDF
Oil, Auto Detect 1 or 2, or Wash Solution
dispense error.
HB
High background in the luminometer.
I
There was no kinetic data for the sample
and neither an NTA nor an NTI flag was
applied to the sample.
LDF
System error.
LPF
The MTU could not be positioned
properly in the luminometer.
M
The sample ID was manually entered
using the keyboard.
MFS
Sample pipetting error.
®
PROCLEIX
TIGRIS System Operator's Manual
Volume I: Operating the Instrument
C
TATUS
ODES
Effect
This is a non-fatal error. The affected
samples are marked Invalid.
This is a fatal error. Every sample on the
worklist is marked Invalid.
The affected samples are marked Invalid.
This error does not count towards the
10% invalid criterion.
This is a non-fatal error. The affected
samples are marked Invalid.
This is a non-fatal error. The affected
samples are marked Invalid.
This is a non-fatal error. The affected
samples are marked Invalid.
This is a non-fatal error. The affected
samples are marked Invalid.
The affected samples are marked Invalid.
All samples in the affected MTU are
marked invalid.
The sample is not marked Invalid.
This is a non-fatal error. The affected
samples are marked Invalid.
S
C
TATUS
ODES
A-1
Part # 902649EN Rev. A