Table 5.5 Hiv-1 Discriminatory Lis File Line Content - Procleix TIGRIS Operator's Manual

Table of Contents

Advertisement

S
5
ECTION
Negative Calibrator
Analyte Average
Negative Calibrator IC
Average
HIV-1 Positive Calibrator
Analyte Average
HIV-1 Positive Calibrator
IC Average
Blank
Blank
Master Lot Number
Master Lot Expiration
Date
Instrument serial #
Firmware Revision
Site Information
Type of tube
Blank
Blank
Test ID
ADM Version Number
TIGRIS Software
Version Number
Example tube data line for a HIV-1 Discriminatory LIS File (comma delimited):
MAS4563,NonReactive,dHIV-1,100376,2005-01-06
21:21:24,IUO,192004,Valid,0,0.00,1.14431,jennifert,93100,57452,326.000,186200.000,1428171.
667,186241.333,,,311348,2025-12-31 23:59:59,HC+1027,HC+E5.11 12/02/04,10120 Genetic
Center Dr,Specimen,,,100376-001311-5-0,2.2.18,2.2.18
All of the information above appears on a single line in the PROCLEIX TIGRIS System LIS file.
5-26
Part # 902649EN Rev. A
Average RLU value for Negative Calibrator
Analyte.
Average RLU value for Negative Calibrator IC.
Average RLU value for HIV-1 Positive Calibrator
Analyte.
Average RLU value for Positive HIV-1 Calibrator
IC.
Reserved.
Reserved.
Master Lot information as entered on the Inventory
Status screen.
Master Lot Expiration Date information as entered
in Inventory Status screen.
Serial # of the instrument, with the prefix "HC+".
Firmware version, with the prefix "HC+".
Identifying site information.
Negative Calibrator, HIV-1 Positive Calibrator,
Specimen, Negative Control, HIV-1 Positive
Control
Reserved.
Reserved.
Test ID in the format: WorklistID-RackBarcode-
PositionInRack-ReplicateNum.
ADM Version Number.
PROCLEIX TIGRIS System Software Version
Number.

Table 5.5 HIV-1 Discriminatory LIS File Line Content

®
PROCLEIX
TIGRIS System Operator's Manual
Volume I: Operating the Instrument
20
20
20
20
0
0
6
19
254 (15 typical)
254 (20 typical)
512
254 (10 typical)
0
0
18
254 (10 typical)
254 (10 typical)