Procleix TIGRIS Operator's Manual page 65

Table of Contents

Advertisement

3. Click OK.
A New Master Lot Components window opens.
4. Scan the remaining barcodes on the Master Lot card for each AR belonging to the Master
Lot. The AR barcodes can be scanned in any order.
5. As each AR barcode is scanned, the barcode populates the appropriate text boxes in the
New Master Lot Components window.
6. Click OK.
Unregistering a Master Lot
Unregistered Master Lots remain in the database, but cannot be used to run assays. To
unregister a Master Lot:
1. Click Unregister Lot... on the Inventory Status screen.
The Unregister Master Lot window opens.
2. Scan the Master Lot barcode on the Master Lot Sheet to enter it into the Barcode field.
3. Click OK.
If the barcode passes the necessary checks, it is unregistered and an Unregister Master
Lot confirmation window opens.
4. Click OK.
Re-registering a Master Lot
To re-register a Master Lot:
1. Click New Lot. A New Master Lot window opens.
2. Using the hand-held barcode scanner on the computer workstation, scan the Master Lot
barcode from the Master Lot Barcode Sheet. A Confirmation window opens, indicating the
Master Lot barcode has been previously unregistered.
3. Click OK to re-register. A final New Master Lot window opens.
4. Click OK to exit.
Resolving an Ambiguous Master Lot
When validating Upper Bay Inventory, "Master Lot is Ambiguous" messages may appear in the
Load Assay Reagents window. This occurs when a reagent is part of more than one Master Lot,
and therefore the system does not know which Master Lot to track the reagent with. The
operator must identify the correct Master Lot in order to accurately track on-board reagent
stability.
To resolve an ambiguous Master Lot:
1. Click the message line to resolve the Master Lot. A "Resolve Ambiguous Master Lot"
message appears.
2. Scan in the correct Master Lot for that quadrant, then click OK.
3. Repeat for any other quadrants that the system identified as having an ambiguous Master
Lot.
4. Click Inventory in the Load Assay Reagents window to continue with Upper Bay
Inventory.
Note: For additional information on resolving ambiguous Master Lots, contact Novartis
Diagnostics Technical Service.
®
PROCLEIX
TIGRIS System Operator's Manual
Volume I: Operating the Instrument
S
I
ETTING UP THE
NSTRUMENT
Part # 902649EN Rev. A
2-21