5. Close the Fluids Drawer and secure it in place by rotating the latch.
Emptying the Liquid Waste
To empty the Liquid Waste containers, first open the waste drawer:
1. Rotate the latch and pull out the drawer.
2. Disconnect the quick-connect fitting on the Liquid Waste Container 2 bottle cap.
3. Park the quick-connect fitting into the slot above the container on the door frame (Figure
2.6) where it will not interfere with Lower Bay maintenance.
4. After the hose is secured, pull the drawer out farther to facilitate access to the Liquid
Waste Containers and MTU Waste Container.
To empty Liquid Waste Container 1:
1. Firmly press the gray plastic vacuum fitting of the left line (leading to the vacuum manifold)
on the bottle cap of Liquid Waste Container 1. When the vacuum releases, pull the hose to
disconnect it from the bottle cap. Apply firm pressure to release the hose fitting.
2. Firmly press the gray plastic vacuum fitting of the right line (leading to the vacuum trap) on
the bottle cap of Liquid Waste Container 1. When the vacuum releases, pull the hose to
disconnect it from the bottle cap. Apply firm pressure to release the hose fitting.
3. Inspect the o-rings on the end of the vacuum hoses to make sure they are present and
undamaged.
2-16
Part # 902649EN Rev. A

Figure 2.6 Liquid Waste Container 2 Fitting

Figure 2.7 Liquid Waste Container 1 Fittings

Left line
Right line
®
PROCLEIX
TIGRIS System Operator's Manual
Volume I: Operating the Instrument