Section 7. Shutting Down; Logging Off; Unloading Ars - Procleix TIGRIS Operator's Manual

Table of Contents

Advertisement

S
7. S
ECTION
Use the Logoff/Shutdown screen to unload Assay Reagents, empty waste containers independent of
Lower Bay Inventory, log off the software, and/or shut down the PROCLEIX

Logging Off

An operator can log out without shutting down the PROCLEIX
another operator will be logging in or if the workstation will not be in use for several hours.
Note: Do not log out of the operator account if the system is going to be shut down. When
the system is shut down, the software automatically logs out the current operator.
To log off, click Logoff Current Operator on the Logoff/Shutdown screen.
The system logs off, removes the username from the Status Bar at the bottom of the screen,
and displays the Logon screen, so that the next operator can logon.
Note: In-process worklists continue processing.

Unloading ARs

ARs must be unloaded before the system is shut down.
Note: The Unload Assay Reagents button is disabled when the system is processing
samples. However, ARs can be unloaded when the system is running an automated
maintenance task.
1. On the Logoff/Shutdown screen, click Unload Assay Reagents....
The System displays a confirmation dialog box warning that the Upper Bay inventory
will be invalid.
2. Click OK.
The AR, TCR, and Sample Doors unlock, and an Unload Assay Reagents window
opens. The window displays a summary of the ARs that are currently loaded.
3. Remove the ARs from the AR Bay.
a) Open the AR Door and remove the AR Carousel Cover with one hand. Still holding
the AR Carousel Cover, remove the AR Carousel with the other hand. Replace the
AR Carousel Cover in the AR Bay.
b) Place the AR Carousel on an absorbent pad.
c) Remove each bottle from the Carousel, and re-cap it with a new cap.
Note: To prevent contamination, always use a new cap when re-capping an AR
bottle.
d) Store the ARs according to the appropriate assay package insert.
e) Remove the AR Carousel Cover from the AR Bay, then replace the AR Carousel.
f)
Place the AR Carousel Cover back on the AR Carousel.
g) Close the AR Door.
®
PROCLEIX
TIGRIS System Operator's Manual
Volume I: Operating the Instrument
D
HUTTING
OWN
S
HUTTING
®
TIGRIS System.
®
TIGRIS System. Log out if
Part # 902649EN Rev. A
D
OWN
7-1