The Assay Components summary lists information for each AR, calibrator, and fluid used in the
processing run:
Master Lot Number and expiration date
AR barcodes and Open/Thaw expiration date
Calibrator barcodes and manufacturer's expiration date
Assay fluids barcodes and manufacturer's expiration date
PROCLEIX
The Test Results Report includes all of the run information about each sample in the worklist:
Test Number
Rack ID and Position
Sample ID – specimen barcode ID or calibrator or control description
Internal Control RLU
Internal Control Results
Analyte RLU
Analyte S/CO
Overall Interpretation
Note: Reactive results are color-coded red; Invalid and Suspect results are color-
coded blue.
Flags
The Reactives Report repeats the information from the Results Report for Reactive samples.
The Invalid Report contains all of the information of the Results Report for any samples that
were marked suspect or invalid by the PROCLEIX
Report also contains legends for any flags, rejected tests, or worklist status messages assigned
to any samples that appear in this report.
®
PROCLEIX
TIGRIS System Operator's Manual
Volume I: Operating the Instrument
®
TIGRIS Controls barcodes and manufacturer's expiration date
W
ORKING WITH
®
TIGRIS System software. The Invalid
R
EPORTS
6-11
Part # 902649EN Rev. A