Defining Site Information

The site-specific information displayed on all PROCLEIX
configurable:
1. On the Configuration tab of the Configuration screen, click Define Site Info....
A Site Information window opens.
2. Enter up to eight lines of site-specific information to be printed on reports, such as the
laboratory name and address.
Note: Each line of text is limited to 60 characters.
3. Click OK to save the changes.
This information will now appear on all future reports.

Barcode Scanner Configuration

The PROCLEIX
barcodes:
Code39
Codabar
Code 128
ISBT 128
During initial configuration of the PROCLEIX TIGRIS System, it is necessary to properly set up
the sample barcode options for the symbologies that are in use at the testing site. This must be
done for both the Sample Carousel Barcode Scanner and the Hand-held Barcode Scanner. The
Sample Carousel Barcode Scanner is configured in the PROCLEIX TIGRIS System Software,
on the Configuration tab. The Hand-held Barcode Scanner is configured using the scanner
itself. The latter step is performed by a Novartis Diagnostics Technical Service representative
upon installation. Any time that an update to the barcode configuration is desired, the
configuration changes must be made for both the Sample Carousel Barcode Scanner and the
Hand-held Barcode Scanner. Otherwise, a configuration mismatch will occur and barcodes may
be interpreted differently using the Sample Carousel Barcode Scanner and Hand-held Barcode
Scanner. Use the instructions below to change the barcode configuration settings.
Caution: Using sample barcodes with check digits is recommended to reduce the risk of
barcode misreads.
Configuring the Sample Carousel Barcode Scanner
To configure barcodes:
1. Click Barcode Configuration....
A Barcode Configuration window opens.
2. Select configuration options to enable interpretation of the barcode symbologies that are
in use at the testing site. Barcode symbologies that are not enabled are reported as
unreadable during sample rack scanning. Do not enable barcode symbologies that are not
in use at the site.
a) Codabar
®
PROCLEIX
TIGRIS System Operator's Manual
Volume I: Operating the Instrument
®
TIGRIS System supports the following barcode symbologies for sample tube
Allow Codabar Barcodes - Enables interpretation of Codabar barcodes.