Specifications; Physical Specifications; Environment; Table 1.6 Size And Weight Specifications - Procleix TIGRIS Operator's Manual

Table of Contents

Advertisement

If applicable, specimens found reactive in the primary assay may be run in individual
Discriminatory Assays to determine if they are reactive for one of the viruses being
assayed or any combination of viruses.
4. After the luminometer reads the amplicon signal, the system dispenses Deactivation Fluid
to destroy any amplicon present in the MTU, aspirates the liquid, and discards the MTU.
The Discriminatory Assay steps are identical to the primary assay steps except that
Discriminatory Probe Reagents specific to a single virus are used in place of the primary assay
Probe Reagent.

Specifications

Physical Specifications

Component
PROCLEIX
System Analyzer
Vacuum Pump
Component
PROCLEIX
Analyzer (Overall space required
is: 117 x 60 x 96 in (297 x 152 x
244 cm, W x D x H)
Computer Workstation

Environment

Ambient Temperature
Humidity
Shock and Vibration
Heat Dissipation
Noise Output
Drain Output
®
PROCLEIX
TIGRIS System Operator's Manual
Volume I: Operating the Instrument
Dimensions (W x D x H)
®
69 x 36 x 72 in
TIGRIS
(175.3 x 91.4 x 183 cm)
17 x 11.8 x 17.4 in
(43 x 30 x 44 cm)

Table 1.6 Size and Weight Specifications

TIGRIS System

Table 1.7 Clearance Specifications

15 to 25° C
20-85%
Compliance with IEC 68-2-6, 7.

Table 1.8 Environmental Requirements

773 W (2637 BTU/Hour) during steady operation
(Conversion is 2637 x .293= 773 W)
Compliant with EN61010-1.
None—Waste fluids drain to removable containers.

Table 1.9 Environmental Output

I
NTRODUCTION AND
Minimum clearance required
24 in (61 cm) vertical clearance
24 in (61 cm) around left side and back
24 in (61 cm) accessible on the right side
36 in (91.4 cm) at front of analyzer
24 in (61 cm) clearance in front
P
D
RODUCT
Weight
1291 pounds (581 kg)
55 pounds (25 kg)
Part # 902649EN Rev. A
ESCRIPTION
1-21