6. If the USB flash memory drive will not be used routinely for saving files, change the
archive path back to the original setting after the files have been saved.
If the files are not successfully saved to the USB flash memory drive, check for the following:
♦
The USB flash memory drive is firmly inserted into the correct USB port
♦
There is enough space on the USB flash memory drive to store the files
♦
The system is in Standby mode
♦
The archive path is properly configured
If these conditions are met but the problem persists, contact Novartis Diagnostics Technical
Service for further instructions.
Saving LIS Files
To save LIS files:
1. Select the worklist to be saved from the Completed Worklist grid view of the Results
screen.
2. If saving to a CD-RW, insert the appropriately formatted disc into the disc drive.
3. Click Transmit to LIS.
4. Check the status of the transmitted files from the LIS State box in the Completed Worklist
grid view to ensure that they are transmitted successfully.
Saving Kinetic Files
To save kinetic files:
1. Select the worklist to be saved from the Completed Worklist grid view of the Results
screen.
2. Click Save Kinetic Data.
The status bar indicates when the kinetic data file has been saved. For example, when
worklist ID 100011 run is saved on March 18, 2004 on instrument serial number 1234,
the status bar will display: Exported Kinetic Data: d:\KIN1234-20040318-112450-
100011.csv.
Saving/Printing Results Report
To save the results report:
1. Click View Results Report on the Worklist Detail grid to view the Results Reports.
2. Click Export.
An Exporting Records window appears. When exporting is complete, it disappears.
To print a Results Report, click Print on the Results Report screen.
Example Filenames
The following are examples of files generated for PROCLEIX ULTRIO
100013, run on instrument serial number 1035 on January 21, 2004 at 1:52 pm:
LIS file: @NA21035-20040121-135234-100013.LIS
Kinetic file: KIN1035-20040121-135234-100013.csv
Result Report: 1035-20040121-135234-100013.pdf
®
PROCLEIX
TIGRIS System Operator's Manual
Volume I: Operating the Instrument
M
R
ANAGING
ESULTS
®
Assay worklist ID
4-7
Part # 902649EN Rev. A