To load the program

For information about the program's softkeys, see the menu descriptions starting on
page 3-51.
• Set the Agilent 35670A Dynamic Signal Analyzer's power switch to off (
then connect the analyzer, test instruments, and printer using GPIB cables.
• If you have the PC Style Keyboard, option 1CL, connect the keyboard to the
analyzer using the keyboard cable (see ''To connect the optional keyboard'' in
chapter 2).
• Insert the Semiautomated Performance Test Disk into the analyzer's disk drive,
then set the power switch to on (
• After the analyzer finishes its power-up calibration routine, press the following
keys:
[
]
Local/GPIB
[
SYSTEM CONTROLLR
[
]
System Utility
[
MEMORY USAGE
[
REMOVE WATERFALL
[
CONFIRM REMOVE
[
]
RETURN
[
]
MORE
[
]
SERVICE TESTS
[
PERFRMANC TEST
• Now go to one of the following procedures to continue:
''To run the program in semiautomated mode''
''To run the program without a printer''
''To run the program in manual mode''
).
l
]
]
]
]
]
Verifying Specifications
To load the program
),
O
3-7