Verifying Specifications
Operation Verification Test Record - Four Channel
Serial Number:_______________________Report Number:____________________
Test Date:___/___/___
Single Ch Phase Accuracy
Measurement
Positive slope, Ch 1
Positive slope, Ch 2
Positive slope, Ch 3
Positive slope, Ch 4
Negative slope, Ch 1
Negative slope, Ch 2
Negative slope, Ch 3
Negative slope, Ch 4
Tach Function (option D01 only)
Measurement
Trigger level +8V Pos
Trigger level +8V Neg
Trigger level −8V Pos
Trigger level −8V Neg
ICP Supply
Measurement
Ch 1 Open Circuit Voltage
Ch 2 Open Circuit Voltage
Ch 3 Open Circuit Voltage
Ch 4 Open Circuit Voltage
Ch 1 Current
Ch 2 Current
Ch 3 Current
Ch 4 Current
Lower Limit
Upper Limit
(deg)
(deg)
4
−4
4
−4
4
−4
4
−4
4
−4
4
−4
4
−4
4
−4
Lower Limit
Upper Limit
(%)
(%)
10
−10
10
−10
10
−10
10
−10
Lower Limit
Upper Limit
26 Vdc
32 Vdc
26 Vdc
32 Vdc
26 Vdc
32 Vdc
26 Vdc
32 Vdc
2.75 mA
5.75 mA
2.75 mA
5.75 mA
2.75 mA
5.75 mA
2.75 mA
5.75 mA
14 of 15
Agilent 35670A
Measured Value
Pass/Fail
(deg)
Measured Value
Pass/Fail
(%)
Measured Value
Pass/Fail
Vdc
Vdc
Vdc
Vdc
mA
mA
mA
mA

Hide quick links: