Agilent Technologies 35670A Service Manual page 189

Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

Agilent 35670A
• Remove the A7 CPU assembly.
See ''To remove CPU'' on page 6-11.
• Set the power switch to on (
• If the power supply LED is still off, set the power switch to off (
reconnect the CPU assembly, and go to Step 6.
Step 5. Repeat the following steps until the assembly causing the Power Supply
G
assembly to shut down is located.
• Set the power switch to off (
• Reconnect one assembly at a time in the following order:
A7 CPU (A7 P10 to A99 J7)
A8 Memory (A8 P1 to A7 J3)
A102 DC-DC Converter (cable to A7 P2)
A11 Keyboard Controller (cable to A7 P1)
A100 Disk Drive (cable to A7 P3)
• Set the power switch to on (
• If the green power supply LED is off, the assembly just reconnected is
probably faulty.
A7 Component Locator, Circuit Side
).
l
).
O
).
l
Troubleshooting the Analyzer
To troubleshoot the power supply
),
O
4-13