Agilent Technologies 35670A Service Manual page 111

Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

Agilent 35670A
Test Configuration Menu Descriptions
Press [
TEST CONFIG
test configuration and the following
softkeys:
[
Agilent 35670A ADDRESS
Prompts you to enter the GPIB address
for the Agilent 35670A Dynamic Signal
Analyzer.
The GPIB addresses equals 100
(interface select code) + (primary
address). The interface select code for
the printer and test equipment is 7 (for
example, if the primary address is 8, the
GPIB address is 708).
[
PRINTER ADDRESS
Prompts you to enter the GPIB address
for the printer. To disable the printer,
set the printer address to 0.
[
]
PROCEDURE
] to display the
Prompts you to select the operation
verification procedure (OP_VERIFY) or
the performance test procedure
(PERFORMAN).
]
[
BEEPER
Toggles the beeper on and off. When
the beeper is on, the program beeps
approximately every 2 minutes while
waiting for you to follow the directions
on the display and press [
[
RETURN
Returns to the ITM_35670A main
menu.
]
[
STOP AFTER
Prompts you to select stop after limit
failure, stop after each measurement, or
do not stop after a limit failure or
measurement. If [
selected, the program stops after the
failing measurement is displayed but
before it is printed. At this point you
can continue on and print the failing
measurement or restart the
measurement.
Verifying Specifications
]
CONTINUE
]
]
] is
Limit Failure
].
3-53

Hide quick links: