Agilent Technologies 35670A Service Manual page 239

Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

Agilent 35670A
• Connect the frequency synthesizer to the analyzer's rear panel EXT TRIG
connector using a BNC cable.
• Press the following keys allowing enough time for the analyzer to trigger
before pressing the next key. Note which trigger modes are failing:
[
]
CHANNEL 1
[
]
CHANNEL 2
[
] (option AY6 only)
CHANNEL 3
[
] (option AY6 only)
CHANNEL 4
[
TRIGGER SETUP
[
CHANNEL LEVEL
200
[
]
mV
[
]
SLOPE NEG
[
ALL CHANNELS
[
CHANNEL DELAY
100
[
]
mS
[
]
0
[
]
S
[
]
Rtn
[
]
ARM SETUP
[
MANUAL ARM
[
]
Rtn
[
]
ARM
[
]
ARM SETUP
[
AUTOMATIC ARM
[
]
Rtn
[
EXTERNAL TRIGGER
[
TRIGGER SETUP
[
EXT LEVEL TTL
[
EXT LEVEL USER
[
USER EXT LEVEL
200
[
]
mV
[
EXT RANGE +/
[
USER EXT LEVEL
0
[
]
V
The message
WAITING FOR mode TRIGGER
triggering.
• Change the frequency synthesizer's amplitude to 0.3 Vp-p.
• Press [
EXT RANGE +/
The analyzer should now trigger.
]
]
]
]
]
]
]
]
]
]
]
− 10 ]
]
2 ].
Troubleshooting the Analyzer
To troubleshoot trigger failures
is displayed when the analyzer is not
4-63