Performance Test
Amplitude Linearity
13 dBVrms
−1 dBVrms
−15 dBVrms
−29 dBVrms
−43 dBVrms
−53 dBVrms
A-Weight Filter
10 Hz
31.62 Hz
100 Hz
1 kHz
10 kHz
25.120 kHz
Channel Match
magnitude
phase
Frequency Accuracy
Anti-Alias Filter
<100 kHz
<1 MHz
Input Coupling
Harmonic Distortion
using HP 339A
using HP 3326A with filter
Intermodulation Distortion
Cross Talk
channel to channel
source to input
Single Channel Phase Accuracy
External Trigger
Tach Function
Input Resistance
ICP Supply
oven circuit voltage
current
Agilent 35670A
Using Recommended Test
Measurement Uncertainty Ratio
±0.00001 dB
±0.01 mdeg
± 0.25 deg †
±132<M > mA
Equipment
>10:1
±0.0020 dB
>10:1
±0.0020 dB
>10:1
±0.0026 dB
>10:1
±0.0046 dB
>10:1
±0.0096 dB
>10:1
±0.0255 dB
0.016 dB
>10:1
0.012 dB
>10:1
0.012 dB
>10:1
0.011 dB
>10:1
0.011 dB
>10:1
0.012 dB
>10:1
>10:1
>10:1
4.8:1
±6.25 ppm
>10:1
± 0.1 dB
>10:1
± 0.3 dB
>10:1
± 0.001 dB
4.46:1
±0.184 dB
10:1
±0.92 dB
10:1
±0.83 dB
>10:1
± 0.1 dB
±1.34 dB
>10:1
280 mVpk
3.6:1 ‡
330 mV
>10:1
±17Ω
17Ω
>10:1
±320 mV
>10:1
Verifying Specifications
Measurement Uncertainty
Using Other Test Equipment
Measurement Uncertainty Ratio
6:1
6:1
3-57