Troubleshooting the Analyzer
To troubleshoot IIC bus failures
Step 4. Check the assemblies on the IIC bus by repeating the following steps for each
G
assembly.
• Set the power switch to off (
• Reconnect one assembly at a time in the following order.
A11 Keyboard Controller (cable to A7 P1)
A10 Rear Panel (cable to A10 P100)
A5 Analog
A1/A2 Input (lower slot)
A2 Input (upper slot)
• Set the power switch to on (
• After the softkey menu appears, (about one minute after power up) press
any key.
• If the analyzer does not respond correctly, the assembly just reconnected is
probably faulty.
Step 5. If the failure still is not isolated, go to page 4-29, ''To troubleshoot fast bus
G
failures.''
4-28
).
O
).
l
Agilent 35670A