General Specifications - Agilent Technologies 35670A Service Manual

Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

Agilent 35670A

General Specifications

Safety Standards
EMI/RFI Standards
Acoustics
Fan speed setting of high, automatic, and off are available. The fan off setting can be enabled
for a short period of time, except at higher ambient temperatures where the fan will stay on.
Environmental operating restrictions Operating:
Ambient temperature
Relative humidity (non-condensing)
minimum
maximum
Vibration (5 - 500 Hz)
Shock
Maximum altitude
AC power
DC power
DC current at 12 V (typical)
Warm-up time
Weight
Dimensions
(Excluding bail handle and impact cover)
IEC 801-3 (Radiated Immunity) Performance degradation may occur at Severity Level 2.
CSA Certified for Electronic Test and
Measurement Equipment per CSA C22.2, No.
231
This product is designed for compliance to:
UL1244, Fourth Edition
IEC 348, Second Edition, 1978
CISPR 11
LpA <55 dB (cooling fan at high speed
setting)
<45 dB (auto speed setting at 25° C)
Operating:
disk in drive
no disk in drive
4° to 45° C
0° to 55° C
20%
15%
80% at 32° C
95% at 40° C
0.6 Grms
2.1 Grms
5 G
5 G
(10 ms 1/2 sine)
(10 ms 1/2 sine)
4600 meters (15,000 feet)
90 Vrms to 264 Vrms (47 to 440 Hz)
350 VA maximum
12 Vdc to 28 Vdc nominal
200 VA maximum
10 A (standard)
12 A (4 channel, option AY6)
15 minutes
15 kg (33 lb) net
29 kg (64 lb) shipping
Height: 190 mm (7.5 in)
Width: 340 mm (13.4 in)
Depth: 465 mm (18.3 in)
Specifications
General Specifications
Storage and
transport
–40° to 70° C
5%
95% at 50° C
3.41 Grms
40 G
(3 ms 1/2 sine)
1-13