Agilent 35670A
• Press the following keys:
[
System Utility
[
]
MORE
[
SERVICE TESTS
[
ADJUSTMTS ]
[ ADC ADJUSTMNT
[
]
OFFSET
Wait for the analyzer to set up the adjustment. The analyzer is ready when the
adjustment message appears on the screen.
• If the oscilloscope display looks like the following figure, go to step 17.
The following describes the signals shown on the oscilloscope display:
'' A'' A straight, horizontal trace in the upper half of the display.
''B'' A "noisy" flat trace at the center of the sine wave trace.
''C'' A clean sine wave in the lower half of the display.
A
B
C
R431 and R405 Correctly Adjusted
• If trace ''B'' is not flat, adjust A5 R431.
• If trace ''C'' is not centered over trace ''B,'' adjust A5 R405.
• Set the power switch to off (
• Disconnect the jumper from A5 TP8 and A5 TP300. Reconnect the cable to
A5 P4.
]
]
]
).