To Set Up The External Trigger Test - Agilent Technologies 35670A Service Manual

Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

Agilent 35670A

To set up the external trigger test

Performance Test only
This test verifies that the Agilent 35670A meets its trigger specification for external
trigger level accuracy. In this test, a frequency synthesizer outputs a 1 kHz signal
to the external trigger input and a 12.8 kHz signal to channel 1. The analyzer
makes an accurate triggered measurement on channel 1 to verify the trigger level
and slope.
1
2 ch
Using an HP 3326A
1
4 ch
Using an HP 3326A
Verifying Specifications
To set up the external trigger test
3-35