Adjusting the Analyzer

To adjust the source

To adjust the source
This procedure adjusts the source dc offset on the A5 Analog assembly.
Equipment Required: Multimeter
BNC-to-Dual Banana Cable
• Connect the multimeter to the analyzer's rear-panel SOURCE connector.
• Press the following keys:
[
]
System Utility
[
]
MORE
[
]
SERVICE TESTS
[
]
ADJUSTMTS
[
SOURCE ADJUSTMNT
[
]
DC OFFSET
• Adjust A5 R48 for 0 Vdc±1 mV.
• Press the [
FILTER DC OFFSET
• Adjust A5 R59 for 0 Vdc ±1 mV.
5-6
]