Agilent Technologies 35670A Service Manual page 122

Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

Verifying Specifications
Performance Test Record - Two Channel
Serial Number:_______________________Report Number:____________________
Test Date:___/___/___
Spurious Signals (continued)
Measurement
One Ch, 88800 Start, Ch 1
One Ch, 97000 Start, Ch 1
One Ch, 98600 Start, Ch 1
One Ch, 100200 Start, Ch 1
One Ch, 101800 Start, Ch 1
Amplitude Accuracy
Measurement
−51 dBVrms, Ch 1
−51 dBVrms, Ch 2
−43 dBVrms, Ch 1
−43 dBVrms, Ch 2
−35 dBVrms, Ch 1
−35 dBVrms, Ch 2
−27 dBVrms, Ch 1
−27 dBVrms, Ch 2
−11 dBVrms, Ch 1
−11 dBVrms, Ch 2
1 dBVrms, Ch 1
1 dBVrms, Ch 2
9 dBVrms, Ch 1
9 dBVrms, Ch 2
19 dBVrms, Ch 1
19 dBVrms, Ch 2
27 dBVrms, Ch 1
27 dBVrms, Ch 2
Flatness
Measurement
27 dBVrms, 99.84 kHz, One Ch, Ch 1
9 dBVrms, 99.84 kHz, One Ch, Ch 1
Lower Limit
Upper Limit
(dBfs)
−80
−80
−80
−80
−80
Lower Limit
Upper Limit
(dBVrms)
(dBVrms)
−51.15
−50.85
−51.15
−50.85
−43.15
−42.85
−43.15
−42.85
−35.15
−34.85
−35.15
−34.85
−27.15
−26.85
−27.15
−26.85
−11.15
−10.85
−11.15
−10.85
0.85
1.15
0.85
1.15
8.85
9.15
8.85
9.15
18.85
19.15
18.85
19.15
26.85
27.15
26.85
27.15
Lower Limit
Upper Limit
(dB)
(dB)
0.2
−0.2
0.2
−0.2
6 of 14
Agilent 35670A
Measured Value
Pass/Fail
(dBfs)
Measured Value
Pass/Fail
(dBVrms)
Measured Value
Pass/Fail
(dB)