Agilent Technologies 35670A Service Manual page 386

Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

Power-on Test Messages
Hexadecimal
Code
Undefined
Initial power-on
FF*
CPU flashes LEDs
04
LED DSACK failure
13
CPU failure
01
Coprocessor DSACK failure
17
Coprocessor failure
18
Boot ROM checksum failure
06
Display DSACK failure
10
Display failure
1B
Main RAM too small
02
Main RAM DSACK
14
Main RAM bit failure
16
Main RAM refresh failure
1C
Program ROM checksum error
00
Clear ~4s
A1
Starting DSP test
A2
Fast bus test
A0
MFP test failure
AE
Front panel test
Internal Test Descriptions
Power-on Test Description
Message
0
Assembly or sub-block is used but is probably not the cause of the failure message.
X
Assembly or sub-block is probably the cause of the failure message.
(blank) Assembly or sub-block is not used in the test.
FF* If the area of failure is unclear, all LEDs flash continuously.
10-4
Assembly/Sub-block
X
X
X
0
X
X
X
0
X
X
X
0
X
X
X
0
0
X
X
0
0
X
X
0
0
0
X
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
X
X
0
0
0
X
0
0
0
0
0
0
0
X
0
0
0
X
0
Agilent 35670A
X
X
0
X
0
X
0
X
0
X
0
X
0
0
0
0
0
X
0
0
0
0
X

Hide quick links: